ATS 2024: Validation of an Acoustic AI Cough Monitoring System

Summary

Over the 477 hours monitored, 4102 cough-seconds were recorded from the 20 subjects; the CoughMonitor detected 3715 of these (overall sensitivity 90.57%) and had 520 false positives (overall false positive rate of 1.09 false positives per hour).