Updated interim data from 54 COPD patients and 135,000 monitored hours, showing cough rates rise in the days before exacerbations and correlate only weakly with CAT scores.

A month of continuous monitoring in stable chronic cough shows that multi-day averaging makes cough frequency a more reliable trial endpoint, allowing smaller studies, and reveals a short-lived Hawthorne effect.

Mixed-methods study in Kampala assessing 48-hour cough frequency monitoring with the Hyfe Research app for TB screening and triage, alongside staff-reported implementation barriers.
24 June 2026

This decentralized study (NCT06775301) validated Hyfe’s smartwatch, which detects and timestamps coughs entirely on-device with no audio recorded or retained. Thirty-one adults wore it for 20 to 24 hours during normal daily activities, while a separate device recorded continuous audio that expert annotators reviewed to mark every cough.
Median sensitivity was 87%, with a median false positive rate of 1.6 per hour. Agreement with ground truth was high: Lin’s concordance correlation coefficient was 0.958 for each patient’s 24-hour total and 0.975 for hourly counts. Performance did not differ significantly by cough rate, sex, age, cough etiology or urban versus rural residence. The device placed 84% of participants in the correct cough-rate category, and the few misclassifications were all minor.